Publication detail
High resolution time-of-flight low energy ion scattering
DRAXLER, M. MARKIN, S. KOLÍBAL, M. PRŮŠA, S. ŠIKOLA, T. BAUER, P.
Original Title
High resolution time-of-flight low energy ion scattering
Type
journal article - other
Language
English
Original Abstract
Low energy ion scattering (LEIS) is a well known technique for quantitative composition and structure analysis. Two different detection methods can be used in a LEIS experiment, i.e. an electrostatic analyzer or a time-of-flight (TOF) spectrometer. Both techniques have specific advantages. Nevertheless, the time-of-flight technique surpasses usual electrostatic analyzers used in LEIS in terms of energy resolution.The possibility to measure ions and neutrals, and the better energy resolution permit to study neutralization or charge exchange processes in much more detail. Here we present a TOF LEIS setup with an energy resolution of better than 0.4% for 3 keV He ions and report experimental results for 3 keV He ions backscattered from a polycrystalline Cu target. The resulting ion spectrum shows interesting inherent features, which are analyzed. Possible causes for the appearance of these features are discussed.
Keywords
High resolution; TOF; LEIS; He; Cu
Authors
DRAXLER, M.; MARKIN, S.; KOLÍBAL, M.; PRŮŠA, S.; ŠIKOLA, T.; BAUER, P.
RIV year
2005
Released
15. 4. 2005
Publisher
Elsevier
ISBN
0168-583X
Periodical
Nuclear Instruments and Methods in Physics Research B
Year of study
230
State
Kingdom of the Netherlands
Pages from
398
Pages to
401
Pages count
4
BibTex
@article{BUT45588,
author="M. {Draxler} and S. N. {Markin} and Miroslav {Kolíbal} and Stanislav {Průša} and Tomáš {Šikola} and P. {Bauer}",
title="High resolution time-of-flight low energy ion scattering",
journal="Nuclear Instruments and Methods in Physics Research B",
year="2005",
volume="230",
number="0",
pages="398--401",
issn="0168-583X"
}