Publication detail
Determination of trace amounts of total dissolved cationic aluminium species in environmental samples by solid phase extraction using manometer-sized titanium dioxide and atomic spectrometry techniques
MATÚŠ, P. HAGAROVÁ, I. BUJDOŠ, M. DIVIŠ, P. KUBOVÁ, J.
Original Title
Determination of trace amounts of total dissolved cationic aluminium species in environmental samples by solid phase extraction using manometer-sized titanium dioxide and atomic spectrometry techniques
Type
journal article in Web of Science
Language
English
Original Abstract
Nanometer-sized titanium dioxide was used as a solid-phase extractant for the separation and preconcentration of trace amounts of AI(III) prior to its determination by electrothermal atomic absorption spectrometry (ET AAS) and inductively coupled plasma optical emission spectrometry (ICP OES). The optimal conditions for the proposed solid phase extraction (SPE; 50 mg TiO(2), 10 min extraction time, pH 6.0, HCl and HNO(3) as eluents) and ET AAS measurement (1500 degrees C pyrolysis and 2600 degrees C atomization temperatures, Mg(NO(3))(2) as matrix modifier) were obtained. The adsorption capacity of TiO(2) was 4.1 mg Al g(-1) TiO(2). Two modes of the proposed procedure were compared, (1) batch and elution mode with the elution of AI from TiO(2) phase by nitric or hydrochloric acid, and (II) batch and slurry mode (without elution) with the direct TiO(2) phase-slurry sampling. Finally, the batch and slurry mode of manometer-sized TiO(2) SPE with slurry ET AAS detection and quantification was preferred and used for the determination of trace amounts of total dissolved cationic Al species in synthetic and natural water samples. The method accuracy was checked by the analysis of lake water CRM TMDA-61 and by the technique of analyte addition (sample spiking). Under the optimal conditions, the calibration curve for batch and slurry TiO(2) SPE with a 10-fold preconcentration was linear up to 40 mu g L(-1) AI. The limit of detection (LOD) and the limit of quantification (LOQ) was 0.11 mu g L(-1) Al and 0.35 mu g L(-1) AI, respectively, with a preconcentration factor of 20 and a relative standard deviation (RSD) lower than 5%. (C) 2009 Elsevier Inc. All rights reserved.
Keywords
Aluminium; Nanometer-sized titanium dioxide TiO(2); Solid phase extraction; Slurry sampling; ET AAS; ICP OES
Authors
MATÚŠ, P.; HAGAROVÁ, I.; BUJDOŠ, M.; DIVIŠ, P.; KUBOVÁ, J.;
RIV year
2009
Released
19. 10. 2009
Publisher
ELSEVIER SCIENCE INC
Location
NEW YORK
ISBN
0162-0134
Periodical
JOURNAL OF INORGANIC BIOCHEMISTRY
Year of study
103
Number
11
State
United States of America
Pages from
1473
Pages to
1479
Pages count
7
URL
BibTex
@article{BUT46848,
author="MATÚŠ, P. and HAGAROVÁ, I. and BUJDOŠ, M. and DIVIŠ, P. and KUBOVÁ, J.",
title="Determination of trace amounts of total dissolved cationic aluminium species in environmental samples by solid phase extraction using manometer-sized titanium dioxide and atomic spectrometry techniques",
journal="JOURNAL OF INORGANIC BIOCHEMISTRY",
year="2009",
volume="103",
number="11",
pages="1473--1479",
doi="10.1016/j.jinorgbio.2009.08.004",
issn="0162-0134",
url="https://www.sciencedirect.com/science/article/pii/S0162013409001652?via%3Dihub"
}