Detail publikace
Oect multielectrode array for real-time cell culture monitoring
SALYK, O. VÍTEČEK, J. ŠAFAŘÍKOVÁ, E. OMASTA, L. STŘÍTESKÝ, S. EHLICH, J. VALA, M. WEITER, M.
Originální název
Oect multielectrode array for real-time cell culture monitoring
Typ
článek ve sborníku ve WoS nebo Scopus
Jazyk
angličtina
Originální abstrakt
The significant cost of newly developed pharmacological products is due to the toxicity examination performed on animals in pre-clinical trials. Our ultimate goal is to develop a biosensor array to determine acute cardiotoxicity. It capitalizes on the fact that heart muscle cells (cardiomyocytes) show beating in vitro. Based on the beating characteristics their pathophysiological status can be deducted. In addition to microscopic imaging, the method of monitoring the action and field potential of synchronously beating cardiomyocyte syncytia on multielectrode arrays (MEA) appears to be a suitable method for the investigation of the condition of cardiac cells. For this purpose, MEA on the basis of organic electrochemical transistors (OECTs) has been developed. The organic polymer PEDOT:PSS was utilized to meet the strict surface biocompatibility demand for a surface in contact with media and living cardiac cells. The MEA consisted of a microplate of a 12x8 chimney-well array with transistors on the bottom for cell cultivation. For the fabrication (a screen printing method) primarily a contact field and an organic polymer was used. The device was tested in a simulation mode for the response of the drain current to the voltage pulses applied to the gate electrode up to 100 mu V. The response to such gate voltage renders the device sensitive enough to detect the electrical signal of beating cardiomyocytes. Further development is directed towards higher time resolution by means of miniaturization of the transistors' channel.
Klíčová slova
microplate; microelectrode array; OECT
Autoři
SALYK, O.; VÍTEČEK, J.; ŠAFAŘÍKOVÁ, E.; OMASTA, L.; STŘÍTESKÝ, S.; EHLICH, J.; VALA, M.; WEITER, M.
Vydáno
17. 9. 2019
Nakladatel
TANGER Ltd.
Místo
Brno
ISBN
978-80-872-9489-5
Kniha
NANOCON 2018 - Conference Proceedings
Strany od
336
Strany do
341
Strany počet
6
URL
BibTex
@inproceedings{BUT163387,
author="Ota {Salyk} and Jan {Víteček} and Eva {Šafaříková} and Lukáš {Omasta} and Stanislav {Stříteský} and Jiří {Ehlich} and Martin {Vala} and Martin {Weiter}",
title="Oect multielectrode array for real-time cell culture monitoring",
booktitle="NANOCON 2018 - Conference Proceedings",
year="2019",
pages="336--341",
publisher="TANGER Ltd.",
address="Brno",
isbn="978-80-872-9489-5",
url="https://www.scopus.com/record/display.uri?eid=2-s2.0-85062984835&origin=resultslist&sort=plf-f&src=s&sid=44723808f725a148aba6e39e5457150f&sot=autdocs&sdt=autdocs&sl=17&s=AU-ID%288378482500%29&relpos=8&citeCnt=0&searchTerm="
}